ST2883 Series Operation Manual Chapter 5 Introduction to [SETUP]
5-2
Start voltage
(Start Volt)
End voltage
(End Volt)
Voltage step
(Volt Step)
Comparator
(Comparator)
Position
(Position)
Limit
(Limit)
State
(State)
5.2.1 Imp Volt
The impulse voltage can be adjusted from 100V to 5000V by a resolution of 10V. Move the cursor to the
impulse voltage zone, following soft keys will be displayed.
They are coarse adjustment keys used to increase and decrease the voltage by a resolution of 100V.
They are fine adjustment keys used to increase and decrease the voltage by a resolution of 10V.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will move
up, down, left and right.
Note: User can directly input the desired voltage by numerical keys. In the state of inputting numbers,
soft keys are unavailable. When the number is input, ENTER can be used to end inputting and the
default unit is volt (V).
5.2.2 Samp
200MSa/s, 100MSa/s, 50MSa/s, 25MSa/s, 12.5MSa/s, 6.25MSa/s, 3.12MSa/s, 1.56MSa/s are optional for
sample rate. Move the cursor to the sample rate zone, following soft keys are available.
Be used to select the desired sample rate.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will move
up, down, left and right.
5.2.3 Test Imp
The number of tested impulses can be set from 1 to 32. Move the cursor to this zone, the following soft keys
will be displayed.
They are fine adjustment soft keys used to change the number of impulses by 1.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will move
up, down, left and right.
Note: User can directly input the desired number by numerical keys. In the state of inputting
numbers, soft keys are unavailable. ENTER can be used to terminate inputting.
5.2.4 Erase Imp
The number of erase impulses can be set from 0 to 16. Move the cursor to this zone, following soft keys will
be displayed.
They are fine adjustment soft keys used to change the number of impulses by 1.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will move
up, down, left and right.
Note: User can directly input the desired number by numerical keys. In the state of inputting
numbers, soft keys are unavailable. ENTER can be used to terminate inputting.