Sorensen MML Series Operation and Programming Manual
7-28
1.8000 represents the load module’s current setting before calibration.
PPPPP represents the load module’s current setting after calibration.
Adjust the knob of the module until shunt current = PPPPP
±
1 count.
Press LOAD and the module displays:
SAvE
Save CCL setting data.
Set the DC source to 5V/20A and current shunt range to 20A.
Press LOAD and the module displays:
9.FS
CCH
Press LOAD and the module displays:
18.000
RRRRR
18.000 represents the load module’s current setting before calibration.
RRRRR represents the load module’s current setting after calibration.
Adjust the knob of the module until shunt current = RRRRR
±
1 count.
Press LOAD and the module displays:
SAvE
Save CCH setting data.
6. CR mode calibration
Set the DC source to 20V/10A and current shunt range to 20A.
Press LOAD and the module displays:
10.FS
CrL
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com