Sorensen MML Series Operation and Programming Manual
4-6
Setting CV Values
Set the CV values for the selected channel by pressing MODE,
^
, and then ENTER
keys. The voltage values can be programmed in one range. The voltage levels are
programmed in volts. The response speed is programmed in fast/slow operations.
The examples below illustrate how to set CV values of the Load module for model
MML-80-60-301. Before following the examples, select channel first.
1. Select
Range
Press MODE and use the
^
or
v
key to select CR followed by ENTER key.
MODE SELECT
CV
2. Set Voltage Level
There are 4000 discrete steps from 0 to full scale in each range. Set the main
voltage level1 (A) to 5 volts by pressing 5, ENTER. Set the level2 (B) voltage level
to 6 volts by pressing 6, then ENTER.
CV 1: 5.00V
CV 2: 6.00V
3. Set Response Speed
There are two response speeds for CV mode, fast and slow, for different UUT
testing. For demanding load tests, use the fast response speed. Refer to
Figures 4-2 and 4-3 for transfer functions.
CV RESPONSE
1:FAST 2:SLOW
Figure 4-2 CV Response Transfer Function (FAST)