e
7-segment Display Unit Indicators. I
ndicate the 7-segment display
measurement unit V and/or I.
f
Operation Mode and GO/NG Indicators.
When the LED of channel R or L is on,
the operation and GO/NG LED have the same function as single channel/module.
When the LEDs of channel R and L are on, the LED of operation mode indicators
will be disabled (off). The GO/NG LED will be red when the check of any channel
SPEC fails. It will be green when the check of both channels SPEC is all right.
g
Keypad Indicators.
There are three LEDs indicating the keypad status. Each
LED shows the key status. It has the same function as single channel/module.
The LED of LOAD will be active when any input of channel L or R is on.
h
Keypad.
There are four keys for you to select/control the operation of Load
module. The R/L key is used to select the display of 7-segment LED, and the
indicators of channel R and/or L. The R/L key can be used to select Fix mode for
rotary knob setting too. Please refer to 4.3.3.
i
Vsense Connectors.
These four connectors are for Vsense measurement input.
The two connectors on the right are for right channel while those on the left for left
channel. Refer to 2.5.2 for remote sensing connections.
j
Rotary Knob.
The knob has the same function as single channel/module when
channel R or L is selected. If the indicators of channel R and L are on, the knob
will be disabled.
k
Load Terminals.
They are input connectors of the Load for connecting to the
UUT. The two terminals on the left are for input of left channel while those on the
right for that of right channel. The PLUS (+) sign of the input of each channel
must connect the high potential. Refer to 2.5.1 for load input connections.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com