9
Caution
When you dispose of the unit or accessories, you must
obey the laws in the relative area or country and the
regulations in the relative hospital regarding
environmental pollution.
Warning on power connections
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each country
if applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power Cord /
Appliance Connector / Plug, please consult a qualified
service personnel.
Warning on power connections for
medical use
Customers in the U.S.A. and Canada should use the
following type of power cord.
Customers in other countries or regions should use the
power cord prescribed by their country or region.
* Note: Grounding reliability can only be achieved when
the equipment is connected to an equivalent receptacle
marked “Hospital Only” or “Hospital Grade”.
Caution
Danger of explosion if battery is incorrectly replaced.
Replace only with the same or equivalent type
recommended by the manufacturer.
When you dispose of the battery, you must obey the law
in the relative area or country.
Warning
When not using the remote for an extended period of
time, remove the batteries to avoid possible damage from
battery leakage and corrosion.
Warning
Batteries shall not be exposed to excessive heat such as
sunshine, fire or the like.
This model (HVO-3300MT) is classified as a CLASS 1
LASER PRODUCT. (IEC 60825-1: 2007 and IEC 60825-
1: 2014)
Caution
Use of controls or adjustments or performance of
procedures other than those specified herein may result in
hazardous radiation exposure.
Caution
The use of optical instruments with this product will
increase eye hazard.
Warning
The apparatus shall not be exposed to dripping or
splashing. No objects filled with liquids, such as vases,
shall be placed on the apparatus.
Caution
When installing, ensure the following space around the
periphery of the unit, taking ventilation and servicing into
consideration.
– Rear side: 10 cm (4 in.) or more
– Left/Right sides: 10 cm (4 in.) or more
– Top side: 10 cm (4 in.) or more
Caution
Do not come into contact with the unit’s internal circuits
and patients at the same time.
Doing so may result in a generation of voltage that can be
harmful to patients if the unit is malfunctioning.
U.S.A. and Canada
Plug type
HOSPITAL GRADE
*
Cord type
Min. Type SJT
Min. 18 AWG
Minimum rating for plug
and appliance couplers
10 A / 125 V
Safety approval
UL Listed and CSA