70
Chapter 8:
Performance Testing
Testing 2D Performance
2D Image Quality
Axial Measurement Accuracy
Note: Measurements must be performed while the image is frozen.
Test 2D
Performance
1
Use a C60e/5-2 MHz transducer in 2D mode.
2
Adjust the position of the C60e/5-2 MHz transducer on the phantom.
3
Use the 2D system controls to obtain a clear image that shows both the horizontal and
vertical rows of pins.
Test 2D Image
Quality
1
Verify that the ultrasound image appears uniform in both the axial and lateral direction,
with no dropouts or intensity variations.
2
Verify that the cystic structure at the focal zone is clearly differentiated from the
surrounding tissue and is echo-free, while solid tissue, with numerous echo sources,
appears solid.
Set up Axial
Accuracy
1
Acquire the image.
2
Press the
Freeze
key.
3
Press the
Caliper
key. The caliper appears on the image display. The screen menu indicates
Cal 1, Cal 2, and Ellipse. (If the caliper line setup is on, then a dotted line connects the two
calipers. See the
MicroMaxx Ultrasound System User Guide
, if necessary.) The Cal 1 caliper is
active by default.
4
Use the touchpad to position one of the calipers.
5
Press the
Select
key to fix the caliper and enable the other caliper.
6
Use the touchpad to move the other caliper. The results update as you move the caliper, and
the measurement is complete when you finish moving the calipers. (Press the
Select
key to
alternate the active caliper, and adjust the measurement with the touchpad.)
7
To perform another distance measurement on the image, select the other distance icon and
repeat the preceding steps.
Test Axial
Accuracy
1
Measure the distance, center to center, of any two pins that are 5-12 cm apart vertically.
2
Verify that the distance measured is within the tolerance listed in
.
Summary of Contents for MicroMaxx
Page 1: ...MicroMaxx Ultrasound System Service Manual ...
Page 6: ...vi ...
Page 10: ...4 Chapter 1 Introduction ...
Page 56: ...50 Chapter 5 Cleaning and Disinfecting ...
Page 61: ...Chapter 6 Troubleshooting 55 Control Panel Figure 6 3 Control Panel Flow Diagram ...
Page 62: ...56 Chapter 6 Troubleshooting System Figure 6 4 System Flow Diagram ...
Page 63: ...Chapter 6 Troubleshooting 57 Battery Figure 6 5 Battery Flow Diagram ...
Page 72: ...66 Chapter 7 Replacement Procedures Figure 7 10 Power Supply Screws Screws 7 ...
Page 85: ...Appendix A Parts List 79 Figure A 1 Power Supply P05471 Figure A 2 Speaker Assembly P03872 ...
Page 89: ...Appendix A Parts List 83 Figure A 8 Main PCB Assembly P05472 ...
Page 92: ...86 Appendix B Service Event Report ...
Page 98: ...92 Index ...