Chapter 7:
Replacement Procedures
67
TGC PCBA
TGC PCBA
1
Remove the TGC knobs identified in
.
Figure 7.11 TGC Knobs
2
Remove the flex cable from the TGC PCB by lifting on the flex release tab. See
3
Remove the flex cable from the Main PCBA by lifting gently on the flex release tab.
4
Remove the two screws holding the TGC PCBA in place.
Figure 7.12 TGC Removal
TGC knobs (3)
Release tabs
Screws (2)