30
Chapter 4:
Setup and Operation
Date and Time
Figure 11 Setup Screen: Date and Time
Ethernet Setup
1
Press the
Setup
key.
2
Select Connectivity.
3
Select Ethernet Setup.
4
Enter information in the following fields: (see DICOM chapter for definition of
terms)
• Host Name
• IP Address
• Subnet Mask
• Default Gateway
• Alternate Gateway
5
In the Network Speed list, select from the list.
6
Select Save.
Reset
Select Reset from the on-screen menu to return settings for this setup page to
factory default.
WARNING:
An accurate date and time are critical for accurate obstetrics calculations. Verify that the date and
time are accurate before each use of the system. The system does not automatically adjust for
daylight savings time changes.
Date
1
Press the
Setup
key.
2
Select Date and Time.
3
In the Date field, enter the current date (year, month, and day).
Time
1
Press the
Setup
key.
2
Select Date and Time.
3
In the Time field, enter the current time in 24 hour format (hours and minutes).
Reset
Select Reset from the on-screen menu to return settings for this setup page to
factory default.
Summary of Contents for MicroMaxx
Page 1: ...MicroMaxx Ultrasound System Service Manual ...
Page 6: ...vi ...
Page 10: ...4 Chapter 1 Introduction ...
Page 56: ...50 Chapter 5 Cleaning and Disinfecting ...
Page 61: ...Chapter 6 Troubleshooting 55 Control Panel Figure 6 3 Control Panel Flow Diagram ...
Page 62: ...56 Chapter 6 Troubleshooting System Figure 6 4 System Flow Diagram ...
Page 63: ...Chapter 6 Troubleshooting 57 Battery Figure 6 5 Battery Flow Diagram ...
Page 72: ...66 Chapter 7 Replacement Procedures Figure 7 10 Power Supply Screws Screws 7 ...
Page 85: ...Appendix A Parts List 79 Figure A 1 Power Supply P05471 Figure A 2 Speaker Assembly P03872 ...
Page 89: ...Appendix A Parts List 83 Figure A 8 Main PCB Assembly P05472 ...
Page 92: ...86 Appendix B Service Event Report ...
Page 98: ...92 Index ...