36
Chapter 4:
Setup and Operation
Presets
Figure 18 Setup Screen: Presets
Edit OB Custom Tables
1
Press the
Setup
key.
2
Select OB Custom Meas. or OB Calculations.
3
Select Tables... from the on-screen menu.
4
Select the desired custom OB table.
5
Select Edit and enter data and then select Save from the on-screen menu.
Delete OB Custom Tables
1
Press the
Setup
key.
2
Select OB Custom Meas. or OB Calculations.
3
Select Tables... from the on-screen menu.
4
Select the desired custom OB table.
5
Select Delete from the on-screen menu to remove the custom table from the
system.
Doppler Scale
1
Press the
Setup
key.
2
Select Presets.
3
In the Doppler Scale list, select cm/s or kHz.
Duplex
1
Press the
Setup
key.
2
Select Presets.
3
In the Duplex list, select desired image display.
• Full 2D, Full Trace
• 1/3 2D, 2/3 Trace
• 1/2 2D, 1/2 Trace
Live Trace
1
Press the
Setup
key.
2
Select Presets.
3
In the Live Trace list, select Peak or Mean.
Thermal Index
1
Press the
Setup
key.
2
Select Presets.
3
In the Thermal Index list, select TIS, TIB, or TIC.
Summary of Contents for MicroMaxx
Page 1: ...MicroMaxx Ultrasound System Service Manual ...
Page 6: ...vi ...
Page 10: ...4 Chapter 1 Introduction ...
Page 56: ...50 Chapter 5 Cleaning and Disinfecting ...
Page 61: ...Chapter 6 Troubleshooting 55 Control Panel Figure 6 3 Control Panel Flow Diagram ...
Page 62: ...56 Chapter 6 Troubleshooting System Figure 6 4 System Flow Diagram ...
Page 63: ...Chapter 6 Troubleshooting 57 Battery Figure 6 5 Battery Flow Diagram ...
Page 72: ...66 Chapter 7 Replacement Procedures Figure 7 10 Power Supply Screws Screws 7 ...
Page 85: ...Appendix A Parts List 79 Figure A 1 Power Supply P05471 Figure A 2 Speaker Assembly P03872 ...
Page 89: ...Appendix A Parts List 83 Figure A 8 Main PCB Assembly P05472 ...
Page 92: ...86 Appendix B Service Event Report ...
Page 98: ...92 Index ...