OPERATING MANUAL PAT-800/805 version 1.09
52
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1%/ºC
Power consumption of
the tested unit
E
4
0%
Low frequency magnetic
field
E
5
0.02 mA I <4mA
0 for I ≥ 4 mA
The shape of the network
voltage (CF)
E
8
0%
measurements are made using mains voltage
bandwidth of current measurement 40Hz ... 100kHz
uncertainty related to the measurement within the bandwidth up to 100kHz should not exceed
± 3dB for 100kHz
adjustable upper limit in the range of: 0.01 ... 9.9 mA with resolution of 0.01mA/0.1mA
adjustable measuring time in the range of: Cont , 4…60s with resolution of 1s
Measurement of differential leakage current
Note: In the half-time of the measurement, the meter automatically changes the polarity of the test
socket and as a final result it displays the value of higher leakage current.
Display range
Resolution
Measurement uncertainty
(basic)
0.00 ... 3.99 mA
0.01 mA
(5% m.v. + 2 digits)
4.0 ... 19.9 mA
0.1 mA
bandwidth of current measurement 20Hz ... 100kHz
uncertainty related to the measurement within the bandwidth up to 100kHz should not exceed
± 3dB for 100kHz
adjustable upper limit in the range of: 0.01 ... 9.9 mA with resolution of 0.01mA/0.1mA
adjustable measuring time in the range of: Cont , 4…60s with resolution of 1s
Influencing factor
Designation
Additional uncertainty
Position
E
1
0%
Supply voltage
E
2
0%
Temperature
E
3
0.1%/ºC
Power consumption of
the tested unit
E
4
Current
common
Additional
uncertainty
0...4A
0
4...8A
±0.03mA
8...16A
±0,08mA
Low frequency
magnetic field
E
5
2 digits I <4mA
0 digit for I ≥ 4 mA
The
shape
of the
network voltage (CF)
E
8
0%