background image

 
 

OPERATING MANUAL PAT-800/805 version 1.09 

1  Safety 

 

 

PAT-800  and  PAT-805  meters  are  designed  for  performing check tests on electrical equipment, 

providing  measurement  results  which  determine  the  safety  status  of  tested  devices.  Therefore,  in 
order  to  provide  conditions  for  correct  operation  and  the  correctness  of  the  obtained  results,  the 
following recommendations must be observed: 

 

Before  you  proceed  to  operate  the  meter,  acquaint  yourself  thoroughly  with  the  present  manual 
and observe the safety regulations and specifications determined by the producer.  

 

Any application that differs from those specified in the present manual may result in a damage to 
the device and constitute a source of danger for the user. 

 

PAT-800  and  PAT-805  meters  must  be  operated  only  by  appropriately  qualified  personnel  with 
relevant  certificates  authorising  the  personnel  to  perform  works  on  electric  systems.  Operating 
the meter by unauthorised personnel may result in damage to the device and constitute a source 
of danger for the user. 

 

The  instrument  must  not  be  used  with  installations  or  equipment  situated  in  dangerous 
environments, e.g. where fire or explosion hazards exist.  

 

It is unacceptable to operate the following: 

  A damaged meter which is completely or partially out of order, 

  A meter with damaged test leads insulation, 

  A meter stored for an excessive period of time in disadvantageous conditions (e.g. excessive 

humidity).  If  the  meter  has  been  transferred  from  a  cool  to  a  warm  environment  with  a  high 
level  of  relative  humidity,  do  not  start  measurements  until  the  meter  is  warmed  up  to  the 
ambient temperature (approximately 30 minutes). 

 

The meter may be supplied only from grounded mains sockets. 

 

Before  measurements  may  commence,  make  sure  the  test  leads  are  connected  to  the 
appropriate measurement sockets. 

 

Do not touch the tested device during measurements. 

 

Banana  test  sockets  and  the  socket  for  testing  IEC  cables  are  protected  against  improper 
connection to the voltage up to 300V AC for 60 seconds.              

 

Repairs may be carried out only by an authorised service point. 

 

 

 

 

NOTE! 

Only standard and additional accessories for a given device should be used, as listed 
in  the "Equipment"  section.  Use  of  different  accessories  can  lead  to  errors in the test 
connection and can introduce additional measurement uncertainties. 
 

 

 

 

NOTE! 

The  plug  on  the  housing  near  handle  must  be  always  tightened.  It  can  be  unscrewed 
only when the device is transported by an aeroplane. 
 

 

 

Note: 

Due to continuous development of the meter’s software, the actual appearance of the 
display, in case of some of the functions, may slightly differ from the display presented 
in this operating manual. 
 

 
 

Summary of Contents for PAT-800

Page 1: ......

Page 2: ......

Page 3: ...OPERATING MANUAL PORTABLE APPLIANCE TESTER PAT 800 and PAT 805 SONEL SA ul Wokulskiego 11 58 100 widnica Poland Version 1 09 13 01 2017...

Page 4: ...09 2 PAT 800 and PAT 805 meters are modern high quality meters easy and safe in operation Please acquaint yourself with the present manual in order to avoid measuring errors and prevent possible prob...

Page 5: ...MEASUREMENT OF THE PROTECTIVE LEAD RESISTANCE USING 200MA CURRENT AUTO ZERO 23 4 4 MEASUREMENT OF PROTECTIVE LEAD RESISTANCE USING 10 25A CURRENT ONLY PAT 805 24 4 4 1 Two wire measurement of protecti...

Page 6: ...5 3 DELETING MEMORY DATA 44 5 3 1 Deleting bank data 44 5 3 2 Deleting the whole memory 45 6 REPORT PRINTING 46 7 DATA TRANSMISSION 46 7 1 COMPUTER CONNECTION ACCESSORIES 46 7 2 DATA TRANSMISSION WITH...

Page 7: ...r which is completely or partially out of order A meter with damaged test leads insulation A meter stored for an excessive period of time in disadvantageous conditions e g excessive humidity If the me...

Page 8: ...g elektrischer Ger te Wiederholungspr fung elektrischer Ger te Allgemeine Anforderungen f r die elektrische Sicherheit Basic functions of the instrument Measurement of network voltage and frequency Ch...

Page 9: ...transfer them to a PC via USB socket or printing cooperation with the bar code reader and printer large readable display with backlight option ergonomic operation Note The displayed E02 symbol inform...

Page 10: ...t When everything is correct the following screen is displayed Notes When the network voltage is below 187V the meter turns off automatically Additional information displayed by the meter and an acous...

Page 11: ...SET button Press ENTER to enter the parameter setting Use buttons to go to the next parameter Use to set the parameter value Press ENTER to approve the setting press ESC go to the main MENU without c...

Page 12: ...9 10 Press ENTER Use to move to the next group of digits using set desired value Press ENTER to confirm the settings or ESC to exit to main MENU without changing any settings 3 3 2 Communication with...

Page 13: ...from www sonel pl Notes New versions of software for the meter are available at www sonel pl This function may be used only by the computer proficient users During programming do not turn off the powe...

Page 14: ...ter codes e g 1234567 any other are considered invalid Therefore if you attempt to read a 6 character code or shorter the reader will read it but PAT will not save it the same applies to 8 character c...

Page 15: ...printer or without it Press ENTER to confirm the settings or ESC to exit to main MENU without changing any settings Note The printer must be connected to any of the USB socket of Host type Supported...

Page 16: ...otes Pen drive must have FAT32 file system Pen drive must be plugged into the left USB socket of Host type The content of memory is transferred to the pen drive as a file in an independent format inte...

Page 17: ...SC to exit to main MENU without changing any settings Notes Nominal network voltage is used in ISUB function for calculating the leakage current which is measured at a voltage of 40V and its value is...

Page 18: ...nd Sonel Reader freeware delivered with the meter enable user to configure meter settings both in terms of general data and individual parameters of each measurement function When using Sonel PAT soft...

Page 19: ...be displayed Window for configuring general settings You can use this module to enter your contact details which will be visible on the reports printed directly from the meter using an optional print...

Page 20: ...asurements Window for configuring Auto measurements The software enables user to read the current configuration of the meter save the meter settings into a file upload the configuration from a file cr...

Page 21: ...is pressed Maintaining the measurement is possible by pressing ENTER with START button kept pressed Each measurement with duration continuous measurement maintaining the measurement by pressing ENTER...

Page 22: ...e is displayed Symbol indicates the possibility to check the fuse removed from the tested device Contact the fuse with the test points The efficiency of the fuse is indicated by displaying and an acou...

Page 23: ...tective lead resistance using 200mA current Press RPE 200mA After that meter is prepared for measurement When parameters must be changed press SET Use to set the upper limit of RPE resistance Use to g...

Page 24: ...of the tested device into the test socket of the meter Use the probe connected to socket I2 for PAT 800 RPE and touch metal parts of the tested device connected to PE Press START After measuring is co...

Page 25: ...ectrically isolated from the mains and from PE mains lead 4 3 Compensation of the test lead resistance during the measurement of the protective lead resistance using 200mA current auto zero In RPE 200...

Page 26: ...ayed message 4 4 Measurement of protective lead resistance using 10 25A current only PAT 805 Press RPE 10 25A After that meter is prepared for measurement Symbol indicates that 10 25 two wire measurem...

Page 27: ...measurement of 200mA Additional information displayed by the meter Non continuity in the test circuit during the measurement 4 4 1 Two wire measurement of protective lead resistance using 10 25A curre...

Page 28: ...g the measurement is signalled by displaying In order to remove auto zero proceed in the same manner but disconnect the test lead from PE After finishing for 1s the message is displayed 4 4 3 Three wi...

Page 29: ...Use the power probe or Kelvin crocodile clip connected to U2 and I2 sockets and touch metal parts of the tested device connected to PE The method of measurement and observations are the same as for t...

Page 30: ...e into the test socket of the meter The measurement is made between shorted L N and PE In addition it is possible to carry out the measurement with the probe connected to RISO socket RISO for PAT 800...

Page 31: ...ould be correct 4 5 2 RISO measurement on devices of Class II III Connect the mains plug of the tested device into the test socket of the meter L and N are shorted Use the probe connected to socket RI...

Page 32: ...UB current is measured at 25 50V voltage and its value is rescaled to the nominal mains voltage set in menu see 3 3 7 Press ISUB After that meter is prepared for measurement When parameters must be ch...

Page 33: ...n Class I additionally connect I2 socket for PAT 800 RISO to the probe which will be used for touching accessible parts of the tested device Press START After measuring is completed read the result Th...

Page 34: ...E mains lead Test voltage is 25 50Vrms 4 7 Measurement of PE leakage current Press IPE After that meter is prepared for measurement Connect the mains plug of the tested device into the test socket of...

Page 35: ...when the device consumes 10A or 16A current Note that if the current leakage is not caused by PE line but by other earthed elements e g water pipe it cannot be measured in this measurement function E...

Page 36: ...rement of a faulty device RCD switch may be triggered off Differential leakage current is measured as a difference between L current and N current This measurement takes into account not only PE leaka...

Page 37: ...nt of touch leakage current Press IT After that meter is prepared fot measurement Supply the tested device from mains socket other than PAT for Class I the socket must have PE Additionally connect the...

Page 38: ...ter that meter is prepared for measurement Press START The measurement may be stopped before the set time runs out by pressing STOP ESC After measuring is completed read the result Symbols and are bli...

Page 39: ...etting is performed as in section 4 2 The following must be successively set LIMIT for RPE measurement time RPE LIMIT for RISO measurement time RISO Connect the mains plug into the test socket and the...

Page 40: ...tomatic tests Press AUTO The screen with test selection will be displayed 1 20 the number of test blinks Notes Only numbers of programmed tests are displayed If no test is programmed after pressing AU...

Page 41: ...t the number of chosen test and confirm by pressing ENTER Use to select the device class and confirm by pressing ENTER Press START to start the measurement To program an inactive test or change the pa...

Page 42: ...asurement or cancel it To change the parameters of the active measurement press SET Set the next parameters as in section 4 2 Confirm by pressing ENTER Use to activate and set parameters of the next m...

Page 43: ...ssing STOP ESC The result remains as at the time of stopping the measurement and the meter turns into standby mode until the next measurement Double pressing of STOP ESC interrupts the cycle of automa...

Page 44: ...mory of measurement result data is not deleted when the meter is switched off Thanks to this the data can be later read or sent to a computer The number of a current memory cell or memory bank is not...

Page 45: ...cupied memory cell the following warning message will appear Press ENTER to overwrite the result or STOP ESC to cancel saving Notes If the meter is set to work with the barcode reader before exiting t...

Page 46: ...ents of the result and other data such as date time of measurement and bar code 5 3 Deleting memory data In the mode of displaying the network voltage press ENTER 5 3 1 Deleting bank data Use and set...

Page 47: ...TING MANUAL PAT 800 805 version 1 09 45 Press ENTER to start deleting or STOP ESC to cancel 5 3 2 Deleting the whole memory Use and set the number of bank at 0 Proceed similarly as when deleting data...

Page 48: ...had the barcode already saved Supported types of printers Brother QL 720NW Brother PT 9700PC 7 Data transmission 7 1 Computer connection accessories In order to operate the meter with a PC an USB cab...

Page 49: ...age In the case of storage of the device the following recommendations must be observed Disconnect all the test leads from the meter Clean the meter and all its accessories thoroughly 10 Dismantling a...

Page 50: ...Hz 2 m v 2 digits measurement of the mains voltage frequency of power supply to the meter Measurement of PE network mains voltage Display range Resolution Measurement uncertainty basic 0 0 59 9 V 0 1...

Page 51: ...nt Influencing factor Designation Additional uncertainty Position E1 0 Supply voltage E2 0 Temperature E3 0 1 C unloaded output voltage 12V AC Test current 10A for R 0 5 adjustable upper limit in the...

Page 52: ...ous voltage before commencing a measurement discharging the object tested Note For R 50k the uncertainty is not specified Measurement of insulation resistance using test voltage of 250V only PAT 805 T...

Page 53: ...ltage before commencing a measurement discharging the object tested Note For R 50k the uncertainty is not specified Measurement of substitute leakage current Display range Resolution Measurement uncer...

Page 54: ...lf time of the measurement the meter automatically changes the polarity of the test socket and as a final result it displays the value of higher leakage current Display range Resolution Measurement un...

Page 55: ...ange Resolution Measurement uncertainty basic 0 999VA 1VA 5 m v 3 digits 1 3 99kVA 0 01 kVA adjustable measuring time in the range of Cont 1 60s with resolution of 1s Current consumption measurement D...

Page 56: ...1kV 1pc WAKROBL30K03 WAKROBL20K01 for PAT 800 1kV probe black 1pc WASONBLOGB3 USB cable WAPRZUSB fuse 0314 015 VXP 15A 250VAC 6 3x32mm Littlefuse 2 pcs WAPOZB15PAT cover WAFUTL5 operating manual 11 3...

Page 57: ...for testing security of devices powered from industrial sockets 16A and 32A providing that the tested device does not consume current higher than 16A The adapters enable users to perform all measureme...

Page 58: ...OPERATING MANUAL PAT 800 805 version 1 09 56...

Page 59: ......

Page 60: ......

Reviews: