The object parameters can be edited. Press on and hold the center part of the object,
without moving it, to open a menu with the object parameters:
The 'Marking' option turns on/off the temperature points displayed on the line. The 'Mark-
ing Content' option allows to select the displayed temperature value for the object: mini-
mum, maximum or average. The 'Center' option moves the object to the center of the
screen.
When the maximum temperature is selected, then the displayed temperature spot for giv-
en object is rendered as a red cross. For the minimum temperature it's rendered as a blue
cross, and for average both crosses are visible.
Remove the line by clicking the
symbol.
7.4 Differential measurement and Delta T mode
It is possible to measure the temperature difference of a two objects of analysis. To do
this, there must be at least two objects of analysis added on screen. Additionally, these
two objects must be set to indicate the temperatures of the same type. It means that when
the first object is set to show the maximum temperature, the second one must also be set
to show the maximum temperature. The camera can compare two maximum, minimum or
average temperatures, never mixed ones. To activate this option press
button on the
pop-up menu: