89
M2 Data Acquisition Device
Updating the Software
The update programming takes about five minutes, then the Home screen opens on the
display device. Create a backup of the VERDICT M2 software to complete the process.
8.3.1 Creating a Software Backup
A backup copy of the newly updated software can be created and stored on the VERDICT M2.
This copy can be used in the unlikely event that the original software becomes corrupted.
z
To backup the software:
1.
Power off the VERDICT M2 unit.
2.
Power on the VERDICT M2 unit and set the Rotary Switch to the
Volts DC
position.
3.
Simultaneously press and hold the
F2
and
S
soft keys for two seconds.
A “set memory recovery” message displays.
4.
Press the Range soft key to initiate the backup.
The screen display returns to the Volts DC screen once the procedure is complete. Both tools
are now ready for use.