HOT LINE
®
Blood and Fluid Warmer
Technical Service Pack
55
S E C T I O N 5
M a i n t e n a n c e S o l u t i o n s
■
Unplug the HOTLINE®.
■
(Old case) Remove the Fill Port plug and hold the Unit over a
sink to empty the water.
(New case) Deploy the drain tube to empty the Unit.
■
Rinse reservoir with distilled water twice.
■
Refill the reservoir with 1.4 litres of disinfectant flushing
solution made to one of the formulae shown in the blue
panel above. Do not fill the HOTLINE® reservoir with a
Disposable Set in place, as this may result in an airlock in the
HOTLINE® warmer.
■
Connect a Disposable Set to the HOTLINE®.
■
Plug the Unit in and switch on.
■
Allow the disinfectant solution to circulate for 30 minutes.
Drain Tube
(in deployed posn)
Drain Tube End Cap
Fill Port
Pole Clamp
Interlock Block
disposable connector
Stowage Clip
(for drain tube when
not deployed)
GM-HL90-8508-en
Figure 5 - 4: Rear view of new-style HOTLINE case, showing drain tube
Summary of Contents for level 1 HOTLINE HL-90
Page 1: ......
Page 14: ...Introduction Specifications Blood and Fluid Warmer ...
Page 20: ...Sub Assemblies Blood and Fluid Warmer ...
Page 28: ...Technical Description Blood and Fluid Warmer ...
Page 45: ...Disassembly Procedures Blood and Fluid Warmer ...
Page 58: ...Blood and Fluid Warmer Maintenance Testing and Calibration ...
Page 83: ...Troubleshooting Blood and Fluid Warmer ...
Page 88: ...Spare Parts Blood and Fluid Warmer ...
Page 96: ...Rounding Off Blood and Fluid Warmer ...
Page 106: ...Appendix 1 Circuits Blood and Fluid Warmer ...
Page 110: ...Appendix 2 PCB Layout Blood and Fluid Warmer ...
Page 115: ...Appendix 3 Symbols Glossary Blood and Fluid Warmer ...
Page 118: ...Appendix 4 Frequently Asked Questions Information Bulletins Blood and Fluid Warmer ...
Page 132: ...Ndjg CdiZh 9 V gVbh ...