HOT LINE
®
Blood and Fluid Warmer
Technical Service Pack
5
S E C T I O N 1
S p e c i f i c a t i o n s
Environmental
Temperature
Operating Range
10
o
C — 45
o
C,
10 — 95% RH, (non-condensing)
Transport / Storage
-20
o
C — 70
o
C
5 — 95% RH, (non-condensing)
Water Resistance
Enclosure Protection
IEC 60529, IP Code: IPX1
Miscellaneous
Standards Compliance
Fluid Warmers
ASTM F2172-02
Product Safety
EN60601-1, UL2601-1
EMC
EN60601-1-2; FCC 47
CFR part 15, Class B
Performance
Recirculating Solution temperature display will reach 37
o
C
from ambient in about 4 minutes.
Operating Temp. Set Point 41.5
o
C ± 0.5
o
C
Over Temp. Set Point
42.6
o
C ± 0.5
o
C
Max Height on IV pole
107 cm (42 inch)
Recirculating Solution
1.4 litres Capacity
Summary of Contents for level 1 HOTLINE HL-90
Page 1: ......
Page 14: ...Introduction Specifications Blood and Fluid Warmer ...
Page 20: ...Sub Assemblies Blood and Fluid Warmer ...
Page 28: ...Technical Description Blood and Fluid Warmer ...
Page 45: ...Disassembly Procedures Blood and Fluid Warmer ...
Page 58: ...Blood and Fluid Warmer Maintenance Testing and Calibration ...
Page 83: ...Troubleshooting Blood and Fluid Warmer ...
Page 88: ...Spare Parts Blood and Fluid Warmer ...
Page 96: ...Rounding Off Blood and Fluid Warmer ...
Page 106: ...Appendix 1 Circuits Blood and Fluid Warmer ...
Page 110: ...Appendix 2 PCB Layout Blood and Fluid Warmer ...
Page 115: ...Appendix 3 Symbols Glossary Blood and Fluid Warmer ...
Page 118: ...Appendix 4 Frequently Asked Questions Information Bulletins Blood and Fluid Warmer ...
Page 132: ...Ndjg CdiZh 9 V gVbh ...