Chapter 5: Troubleshooting
5-2
SurgiVet
®
Universal CDS 9000 Operation Manual
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Rebreathing bag over
inflates (Pressure is over 5
cm H
2
O)
Size of bag is inappropriate for the size
patient it is being used on.
Pop-off valve is partially or completely
closed or not functioning.
Flow is obstructed passing through the
Charcoal Canister.
Flush Valve is defective.
Check the size of the bag and change if
needed.
Open Pop-off valve all the way or Clean
or replace Pop-off valve.
Position charcoal canister so gas can pass
through properly.
Check and replace Flush Valve if
necessary.
If the SurgiVet
®
CDS 9000 Anesthesia Machine is still not functioning properly, contact Veterinary Clinical
Support for help.