Chapter 3: Machine Operation
3-2
Surgivet
®
Universal CDS 9000 Operation Manual
To setup the machine:
Gather all parts from the shipping container.
1.
Firmly insert casters into the base to ensure a balanced base (See Figure 3-1).
2.
Insert the beveled (tapered) end of the 2” Aluminum Pole into the Base and press firmly to ensure stability.
3.
All pole mounted optional accessories for the anesthesia machine should be installed at this time. Slide
4.
each accessory down over the pole and secure it in position before proceeding to the next step. Optional
Accessories are as follows:
V7330 O2 E-Tank Regulator, Double Tank Holder
•
V7331 Universal Double E-Tank Monifold
•
V7261 E-Tank Stabilizer
•
V7115 Mayo Tray
•
V7260 SAV Universal Mounting Bracket
•
V7121 “Z” Bracket
•
V7122 Universal Vaporizer Mounting Bracket with end caps
•
V712201 Universal Vaporizer Mounting Bracket without end caps
•
Align the Pole Plug with the hole in the 2” aluminum pole. Install the machine head onto the pole. Install
5.
the 1/4 - 20 x 1 1/2 Button Head screw and tighten.
Cut the plastic zip tie and carefully release the chrome end caps with silicone tubing from the anesthesia
6.
machine bracket.
Attach the vaporizer to the anesthesia machine bracket using the screws provided (located on the back of
7.
the vaporizer manifold).
NOTE! Vaporizer mounting hardware is included with the purchase of the vaporizer only; it is not
included with the purchase of the anesthesia machine due to the various styles of vaporizers that
the machine accommodates.
NOTE! The order of the vaporizer mounting hardware on the bolt is as follows:
Lock Washer---Flat Washer---Bracket (machine) ---Nylon Washer---Vaporizer
NOTE! The nylon washer should be placed between the vaporizer manifold and the frame of the
machine. This allows enough space to appropriately attach the male and female endcaps.
NOTE! It is not necessary to use a Nylon Washer for the Tech 4 Vaporizers
NOTE! For certain styles of vaporizers, it may be necessary to attach the male and female end caps to the
vaporizer first, before mounting it to the machine (see step 8).
If you have not done so already, twist and pull the end caps apart and attach the end caps to the manifold
8.
of the vaporizer using a push and twist action. The male end cap goes on the outlet of the vaporizer (right
side when looking at the front of the vaporizer) and the female end cap goes on the inlet of the vaporizer
(left side when looking at the front of the vaporizer).