Chapter 4: Routine Maintenance
Surgivet
®
Universal CDS 9000 Operation Manual
4-9
Once Flowtube and ball are dry, install ball and stopper. Make sure stopper is slightly recessed for alignment
12.
guide.
Figure 4-14: Recessed Stop
Install Flowtube assembly. Make sure Flowtube is centered on the alignment guide located inside the
13.
gaskets. (Top and Bottom)
Figure 4-15: Stainless Steal Alignment Guide
While holding the Flowtube the entire time, center the Flowtube and tighten screw with 5/32” Allen Wrench.
14.
Make sure numbers are facing out toward you. Do not overtighten.
Replace plastic cover.
15.
Turn oxygen on. Test Flowmeter by turning it fully On and then fully Off gradually. If the ball continues to
16.
stick in the tube, repeat steps 1-16 or contact Clinical Support.
Recessed Stop
Stainless Steal Alignment Guide