![Smiths Medical CDS 9000 Operation Manual Download Page 28](http://html.mh-extra.com/html/smiths-medical/cds-9000/cds-9000_operation-manual_1306035028.webp)
Chapter 4: Routine Maintenance
4-4
Surgivet
®
Universal CDS 9000 Operation Manual
Pop-off Valve Removal and Cleaning
Figure 4-1: Pop-Off Valve Removal (Supplies needed)
NOTE! Pop-Off Valve shown disassembled from the machine for clarity. It is not necessary to disassemble
the Pop-Off Valve from the machine to perform this operation.
Supplies Needed:
5/64 Allen Wrench
•
70% Isopropyl Alcohol - not shown
•
Soft Cloth
•
Krytox® Lubricant
•
Needle-Nosed Pliers - not shown
•
Loosen the small set screw located on the chrome body of the Pop-off Valve. (See Figure 4-2)
1.
Open the Pop-Off Valve to the fully open position (counterclockwise) and then turn 1/4 turn back
2.
(clockwise). Unscrew the top black portion of the Pop-off Valve and then unscrew the assembly from the
body. (See Figure 4-3)
Figure 4-2: Loosen Set Screw on Pop-Off Valve
Figure 4-3: Unscrew Assembly from Body