Chapter 4: Routine Maintenance
Surgivet
®
Universal CDS 9000 Operation Manual
4-1
Chapter 4: Routine Maintenance
Routine Maintenance
ITEM
ACTION
INTER-
VAL
PART NUMBER
Pressure Test
See Daily Machine Checklist
Daily
NA
Patient circuits, hoses,
rebreathing bags
Clean, inspect for damage
Daily
Universal F-circuit - 32359B3
Modified Jackson Rees - 32359B6
Machine surfaces
Clean
Daily
NA
CO
2
absorber canister Clean. Change when 2/3 blue/purple As needed Absorber canister-32046B1
Inhalation/Exhalation
Valves
Visually inspect
Daily
Dome cap-32166B1
Pop-off Valve
Ensure smooth opening/closing, can
be cleaned
Daily
V7306
Gaskets (absorber and
dome cap)
Visually inspect, clean
Weekly
Absorber gasket-32380B4
Dome cap gasket-32405B16
Flutter disks
Visually inspect for damage/curling,
can be replaced
Weekly
32369B3
Manometer
Zero adjustment
As needed 32027B1
Flowmeter and
Needle Valve
Ensure smooth flow and complete
shut off. (Can be cleaned or replaced
as needed)
Daily
Needle Valve-32413B2
Waste Gas Scavenging System (not included)
Following the manufacturer’s recommendations, check that the scavenging system is properly connected,
turned on and does not have too much draw (bag will be completely deflated if there is too much draw). Visually
inspect all connections to ensure waste gas is not leaking into the operating environment.
WARNING! Waste Gas Anesthetic levels beyond 2 parts per million can cause health hazards.
Pressure Manometer
Visually inspect for proper alignment. Needle should be at the 0 mark. To adjust the manometer back to Zero,
remove the plastic lens from the front of the manometer with a small flat headed screwdriver. There is an
indentation in the cover to aid in removal. Adjust the needle using the screw located directly above and center
on the zero mark.
Long Term Storage
Storage Facility:
Indoor
Temperature:
-1 to +48° C (30 to 120° F)
Relative Humidity:
10-95%, non-condensing
Periodic Inspection:
None required.
Special Procedures:
Store the SurgiVet
®
Universal CDS 9000 Anesthesia Machine in the original packing
materials and shipping carton.