Chapter 2: Description of Controls and Features
2-2
Surgivet
®
Universal CDS 9000 Operation Manual
Items Included With WWV9010 Series
Universal CDS 9010 Table Top
•
32816B1 Draw Knob
•
32046B1 Absorber Canister
•
32359B3 F-Circuit with 2 Liter Bag *
•
32359B6 Modified Jackson Rees with ½ Liter Bag *
•
V7309 Pop-Off Occlusion Valve
•
32046A7 CO
•
2
Absorbent Refill Bag (2.2 lbs) *
Operation Manual *
•
(Items with * are not pictured)
NOTE! Items included with the WWV9010 series are similar to items shown in Figure 2-1.
Definition of parts
Flowmeter
Located to the left of the Vaporizer and Absorber System. The Flowmeter controls the delivery of oxygen into
the breathing circuit. The standard Flowmeter reads from 200cc to 4 Liters Per Minute (other combinations and
ranges are available).
Flush Valve
Located above the Flowmeter, this device is used to quickly deliver oxygen to the patient (bypasses the
Vaporizer). It is also used to pressure test the anesthesia machine.
Female End Cap
Universal (23 mm taper) connection for most vaporizers, attaches to the Vaporizer Inlet.
Male End Cap
Universal (23 mm taper) connection for most vaporizers, attaches to the Vaporizer Outlet.
Pop-Off Valve
Located behind the Pressure Manometer. The Pop-Off Valve directs the waste anesthetic gas from the machine to
a scavenging system. In the completely open position, the SurgiVet
®
Pop- Off Valve will release pressure at 2 cm
H
2
O, while maintaining a constant passive volume in the breathing bag.
NOTE! Do not overtighten the valve when closing.
Pressure Manometer
Located on top of the machine behind the exhalation valve. The manometer has two functions:
Used during a pressure test of the anesthesia machine for leak detection.
•
Provides a visual measurement of the patient airway pressure.
•