Chapter 4: Routine Maintenance
4-6
Surgivet
®
Universal CDS 9000 Operation Manual
Loosen set screws (3) on the black base using a 5/64 inch Allen Wrench. (See Figure 4-5)
15.
CAUTION! Do not completely remove the screws, doing so will expose ball bearings which can be lost
easily.
Slowly remove Chrome Body from Black Base. Be careful not to remove the 3 ball bearings located behind
16.
the set screws. (See Figure 4-6)
Replace the existing o-ring with a new o-ring (P/N 32405B18).
17.
Figure 4-7: Replace O-Ring
Apply a thin film of Krytox to the new o-ring.
18.
Install chrome body back into the black base.
19.
Position Pop-Off as desired and tighten the 3 set screws using a 5/64 inch Allen Wrench.
20.
Perform a pressure test to confirm that the o-ring is sealing properly. Contact Clinical Support if you have
21.
any questions.
Figure 4-5: Loosen Set Screws
Figure 4-6: Remove Chrome Body from Black Base