![SmartLine RotaScope Classic Series Installation And User Instructions Manual Download Page 82](http://html1.mh-extra.com/html/smartline/rotascope-classic-series/rotascope-classic-series_installation-and-user-instructions-manual_1297791082.webp)
RotaScope Classic Cabinet
Manual Version
1.04
Date
23 November 2021
82
4.2.10
Required Tools
Never attempt to lift or move a cabinet without the appropriate training and support,
always use two technicians to move and lift a cabinet.
Required tools:
•
Battery Drill/Driver (Philips Head No2 Driver)
•
Level
•
Shifter (adjustable wrench)
•
Socket set
•
Box Cutter (Stanley Knife or similar)
•
Platform Ladder
•
Dolly Trolley
•
Screwdriver set
•
Wall plugs (Dry wall anchors - Wall Mates)
•
Stud Finder
•
Hospital Grade Disinfectant
4.3
How to Commission the Product
Prior to use, the cabinet must undergo thorough disinfection firstly by the Approved
Distributor and finally at a clinical level clean should be performed by the
Hospital/Departmental Staff.
In-service training will be carried out by an appropriate training provider and is
supplemented by video training.
4.3.1
Installation of the software
•
All cabinets are provided with pre-installed software.
•
See also Communication Kit
–
If supplied, the laptop computer is
loaded with the Scope Manager Software and is used for basic setup
of Operator ID Barcodes and Scope ID Barcodes.
•
Scope Manager Software may be supplied by the Authorised
Distributor and may be installed onto a hospital provided PC under the
supervision of the Authorised Distributor
4.3.2
Calibration of the device
Calibration of the cabinet will be done firstly inhouse at the commencement of
manufacture and is checked again during the Install and Operational Qualification
(IQ,OQ) by the Authorised Distributor.