2.11
Explanation of Auditory and Visual Signals on Cabinet System
2.11.1
Audio Signal at Cabinet Control Screen Loading and Unloading Scopes
Audio Signals at Cabinet Control Screen
No Icon
Description
Location Tone
1.
Scan Operator ID Barcode (Staff ID Badge) -
Unlocks door
Audio Tone -
“
Operator
”
Door
Control
Screen
Voice
2.
Scan Scope ID Barcode
Audio Tone -
“
Scope
”
(See SmartStation section below)
(Scope must
have been scanned out of an AER
–
i.e. Clean)
Door
Control
Screen
Voice
3.
Cabinet Cleaning - Scan Operator ID Barcode
(Staff ID Badge) - Audio Tone -
“Operator”
- then
press Cabinet Clean Button on Main Screen
Audio Tone -
“Cleaned”
Used when cleaning cabinet weekly
Door
Control
Screen
Voice
4.
Loading a Scope
–
Scan Operator ID Barcode
(Staff ID Badge) Audio Tone -
“Operator”
Scan Scope ID Barcode - Audio Tone -
“Scope”
Open Door - Load Scope and Connect Scope
Hose Connection Kits
–
Close Door - Audio Tone -
“
Scope, Stored, Location
X
”
Door
Control
Screen
Voice
5.
Unloading Scope
–
Scan Operator ID Barcode
(Staff ID Badge) Audio Tone -
“Operator”
–
Disconnect and remove scope
–
Close Door
–
Scan Scope ID Barcode
Audio Tone -
“Scope, Location X, Removed”
Door
Control
Screen
Voice