RotaScope Classic Cabinet
Manual Version
1.04
Date
23 November 2021
46
operation, if an operational parameter exceeds a set operational value an Alarm or
Warning may be triggered, and the following will occur:
•
An automatically generated printout identifies the issue (where applicable)
•
Screen display will turn red to highlight issues
•
Cabinet lighting will change colour (each fault has an assigned colour)
Read and understand this manual and its safety instructions before using this
product. Failure to do so can result in serious injury or death.
3.1
How to Use the Product Safely
•
Some staff may have difficulty reaching scopes when loading or unloading the
cabinet, due to the height of the loaded scopes. It is advisable that an
appropriate step stool be used. Consult with your local Occupational Health
Officer to determine a safe work method for your site.
3.1.1
Control screens, and electrical enclosures
•
Do not open and access the control units without specific training from the
Authorised Distributor.
3.1.2
Technical life span
•
Annual service and preventative maintenance are required
o
To ensure you get the best performance out of your Cabinets, it is an
operational requirement to have a Maintenance Contract in place.
These contracts are to be set up through the Maintenance Department.
A Maintenance Contract is an annual service agreement covering
replacement equipment e.g., filters and parts for the Integrated Air
System IAS, and basic ongoing cabinet maintenance of cabinet fittings
and fixtures.
•
The viable lifecycle of the cabinet is 7 years from date of purchase
3.1.3
PQ Performance Qualification
•
Annual Performance Qualification required. EN16442:2015 states in section
E.2
and Annex D - Performance qualification and routine testing
–
must be
performed