RotaScope Classic Cabinet
Manual Version
1.04
Date
23 November 2021
12
2
Description of the product
2.1
Intended Use and Reasonably Foreseeable Misuse
This RotaScope has been designed to dry and store up to nine Flexible Thermolabile Endoscopes.
The RotaScope is intended to be used for Operators in a healthcare environment to store and dry
Thermolabile Endoscopes after being reprocessed.
The RotaScope shall not be used for any other function except for the storing and drying of
Thermolabile Endoscopes.
The RotaScope shall be used with the following original accessories and components only:
•
SmartStations
•
Communications Kit (including the export version
–
excluding laptop)
•
Endoscope Hang Hooks
•
Optional - Approved Endoscope Hose Connection Kits (Purchased Separately)
•
ScopeManager software application
It is essential that the user of this manual is aware of the potential hazards associated with the unit
and its accessories. All operators should be familiar with the safety precautions and warnings given
in this section prior to attempting to operate the unit. If the unit is used in a manner not specified by
the manufacturer, the protection provided by the equipment may be impaired.
2.2
Process Overview
The RotaScope Data System functions on a Linux system and Barcode Scanners. The system functions
on the concept of WHO, WHAT and WHERE, with WHEN being automatically captured by the Data
System.
•
WHO = Operator ID Barcode
•
WHAT = Scope ID Barcode
•
WHERE = Device/Function Barcode
AER Validation:
•
WHO - Operator ID
–
the person that washed the scope in the AER
•
WHAT - Which Scope
•
WHERE - Where the scope was washed (Washer/AER Chamber)
•
The Scope is loaded into the AER