The following filling options can be set for the dosing mode
Fig. 108
If”off“is selected for filling, filling it will not occur automatically after each dosing step.
If “intelligent before“ is selected for filling, a verification will be performed each time prior to the next
dosing step in order to determine whether the dosing step can still be made without a filling operation.
Should this prove to be impossible, the first thing to occur is filling, followed by the dosing step.
If “intelligent after“ is selected for filling, a verification will be performed after the next dosing step to
find out whether the next dosing step can still be made without filling.
If „always“ is selected for filling, filling will occur automatically after each dosing step.
4.6.7
Sample identification
In the manual titration and in the preparation of solutions it is possible to input a sample identification.
The possible input includes manual, automatic or no sample description at all.
Fig. 109
For a sample description of the ’manual’, a prompt for the sample description will always be displayed
at the start of the method (Cp. also chapter 3.6, Main menu). For an ‘automatic’ sample description
there will be selected a master description (in the current case this is water, cp. Fig. 110), which will
then automatically be numbered starting on 01.
Fig. 110
After a new power-up, numbering will resume with 01.