121
<Measuring speed>
or drift will determine the span of time after which the measured value will be
accepted following a titration step.
Fig. 89
Drift-controlled acceptance
of the measured value in terms of mV/min is set by selecting “
norma
l“,
“fast or
”user-defined”
. The drift values at predefined in terms of in mV/min for normal and fast drift:
Normal drift
20 mV/min
Fast drift
50 mV/min
Small drift value = slow and precise
Large drift value = fast and “less precise”
The following parameter selection can be made for user-defined drift setting:
Minimum holding time [s]:
01 - 99
Maximum holding time [s]:
01 - 99
Measuring time t: [s]
01 - 99
Drift [mv/min]
01 - 99
Fig. 90
If normal or fast drift was selected before, the values will be defaulted for user-defined drift. In the
present case, for instance, 20 mV for normal drift: