Form 40017 Rev 1906

Page 12 of 26

www.skcinc.com

To set Secure Lock to on or off and set passcode
From Main Menu: 

 

Touch Device 

> Touch Screen 

> Touch 

Secure Lock

Secure Lock helps to prevent tampering during a sample run. Touch desired bu

tt

 on. 

If Secure Lock was set previously to On and you select 

O

 ,

 you will be prompted to enter the previously set passcode. Once the passcode is entered, you will be returned to the Screen Menu.

If you select 

On

, proceed as follows.

You will be prompted to enter a four-digit passcode of your choosing. Touch the screen keypad to 
enter the desired four-digit number combination. 

Note:

 Entered digits will display as ****.

You will be prompted to con

fi

 rm the passcode. Touch the screen keypad to enter the same four-

digit number combination. Upon entering the last digit, the passcode will be saved and you will 
be returned to the Screen Menu.

If Secure Lock is set to 

On

, the user can lock the screen of a running pump by pressing the Power 

On/O

  bu

tt

 on. If Auto Lock is set to On concurrently, the screen locks when the pump is run.

To unlock and reactivate the screen,

 press the Power On/O

  bu

tt

 on on top of the running pump 

and touch the screen keypad to enter the previously set four-digit passcode. 

Master Unlock Feature:

 If you cannot remember the Secure Lock passcode, touch 

1 2 3

 

 

 

 

 

when prompted for the passcode. 

This will override the Secure Lock, but will not disable it.

Summary of Contents for Pocket Pump TOUCH

Page 1: ...fying Device Settings Clock Menu Date Time Units Menu Display Units Screen Menu Dim Security Operation Sample Menu Set Flow Calibrate and Sample 13 To set pump flow rate To set up a single sorbent tube train and calibrate Constant Flow Mode To set up a multiple sorbent tube train and calibrate Constant Flow Mode To set up a multiple sorbent tube train and calibrate Constant Pressure Mode To calibr...

Page 2: ...e subject to special shipping regulations Single kits are also available in hard sided Pelican case Contact SKC Quick Guide Charging the battery Insert USB connector on cable into USB port on supplied wall adapter Plug into 100 to 240 V outlet Hover magnetic charging connector on cable closely over contact on back of pump until engaged and LEDs on contact and cable light blue Full charge is achiev...

Page 3: ...ging contact but the charger and charging contact LEDs are not lit blue Check that the charging cable is connected to an AC adapter and active outlet PC laptop connected to an active power source or a portable power pack with a full charge that is compatible with charging Pocket Pump TOUCH If only the charger LED lights blue It means that there is no connection with the pump charging contact Check...

Page 4: ...laced in an area to be known nonhazardous Afin de prévenir l inflammation d atmosphères dangereuses ne changer les batteries que dans des emplacements désignés non dangereux Maximum charge input voltage Um 5 5 VDC The battery used in this device may present a risk of fire or explosion when heated above 212 F 100 C or incinerated Replace battery with SKC Battery Pack Cat No P76303 only Use of anoth...

Page 5: ...dicated below Navigating Menus and Screens Pocket Pump TOUCH operates through a series of menus and screens When the pump is powered on the Flow screen displays which allows you fast access to calibration and sampling For more details start at To set pump flow rate Immediately below the display are four virtual navigation buttons that access previous screen Main Menu and increase decrease values B...

Page 6: ...d to pump Advanced Menu Set Pressure Set Duration timer Run button Menus and screens can contain the following navigational touch buttons Button General Function Check mark Saves a selected item Left and right movement Allows horizontal movement on a scale see right or moves left or right through fields activating each for entry of a value Calibration Flow Adjustment Allows selection of an adjustm...

Page 7: ... on the back of the pump These indicate pump status at a quick glance LED Activity Description Pump Status Corner and inlet LEDs green flashing Pump running Corner and inlet LEDs red flashing Pump is in flow fault Inlet LED yellow orange flashing corner LEDs green blinking Pump in constant pressure mode working to achieve pressure setting Inlet LED will change to solid green when set pressure is a...

Page 8: ...ce Touch Clock Touch time displayed Hour digits will flash Touch up down arrow buttons to increment decrement hours Touch right arrow to advance to minutes will flash and up down arrow buttons to adjust minutes Touch right arrow to advance toAM PM will flash and up down arrow buttons to toggleAM PM Touch check mark to accept new time and return to Clock Menu New time setting will display To change...

Page 9: ...ll flash and up down arrow buttons to increment decrement day Touch right arrow to advance to year will flash and up down arrow buttons to increment decrement year Touch check mark to accept new date and return to Clock Menu New date will display To change date display format From Main Menu Touch Device Touch Clock Touch Date Display Format Touch the desired date display format to select and retur...

Page 10: ...hange atmospheric pressure display units From Main Menu Touch Device Touch Units Touch ATM mbar inHg and mmHg buttons display Touch desired button to select and return to Units Menu Note Changing the display units affects only the display of atmospheric pressure on the pump screen and in Sample Summary and History Back pressure inlet pressure will always display in inH2O on the pump screen and in S...

Page 11: ... screen backlight will stay on during the entire sample run To set Auto Lock to on or off From Main Menu Touch Device Touch Screen Touch Auto Lock Touch desired button to select and return to Screen Menu If Auto Lock is set to On the screen will lock become inactive as soon as the pump starts running a sample A lock icon and Press top button to unlock message will appear on the screen If Dim is se...

Page 12: ...combination Note Entered digits will display as You will be prompted to confirm the passcode Touch the screen keypad to enter the same four digit number combination Upon entering the last digit the passcode will be saved and you will be returned to the Screen Menu If Secure Lock is set to On the user can lock the screen of a running pump by pressing the Power On Off button If Auto Lock is set to On...

Page 13: ...rbent tube and insert arrow printed on tube pointing toward the pump into the rubber sleeve of a 222 Series non adjustable Low Flow Tube Holder Figure 3 2 Connect tubing from the 222 Series non adjustable low flow tube holder to the pump inlet Figure 3 3 Use a length of flexible tubing to connect the exposed open end of the representative sorbent tube in the holder to a calibrator outlet Figure 3 ...

Page 14: ...rough all tubes to be used during the multiple tube sample run to ensure adequate airflow Note The combined flow through all tubes cannot exceed 500 ml min 2 Press tubing from the outlet unlabeled side of a Constant Pressure Controller CPC onto the pump inlet Figure 4 3 Press tubing from the 224 Series Multiple tube Low Flow Holder onto the CPC inlet labeled To Sample Figure 4 4 Break tips off a re...

Page 15: ...s on holder operation It is good practice to run the pump for 5 minutes before calibrating the flow rate Calibrate verify pump flow rate before and after each sampling operation using the tube holder and pump to be used for sampling 1 Determine back pressure to be set in Constant Pressure Mode a Connect tubing from holder outlet to pump inlet Figure 5 b Break tips off a representative sorbent tube ...

Page 16: ...to turn flow adjust screw on port clockwise to decrease counterclockwise to increase until calibrator displays the method specified flow rate Note The flow rate displayed on the calibrator is the flow rate through the sorbent tube 6 Repeat Steps 3 through 5 for all remaining open sorbent tubes 7 Disconnect representative sorbent tube from the calibrator Replace representative sorbent tubes with ne...

Page 17: ... Flow Touch calibration icon Pump will run Allow pump to run for 5 minutes before performing calibration Touch up down arrow buttons to increment decrement calibration adjustment A calibration adjustment value will display as shown below left Note The flow rate displayed on the calibrator will change as a result of this adjustment The flow rate displayed on the pump will not change Touch check mar...

Page 18: ...ag sampling a Use PTFE tubing to connect pump exhaust Figure 1 to the bag fitting See bag operating instructions b Unscrew cap on pump exhaust to remove it and screw in supplied exhaust fitting or optional Quick connect Bag Sampling Adapter accessory c Attach tubing from the bag to the fitting or adapter on the pump exhaust 3 On the pump choose from a manual sample timed sample or sample preset pr...

Page 19: ...o Lock is set to Off the screen remains active The screen may be locked made inactive at any time during sampling by pressing the Power On Off button on top of the pump To unlock and reactivate the screen press the Power On Off button on top of the pump This feature helps to reduce tap errors during sample runs Secure Lock set to On locks inactivates the pump touch screen when the Power On Off but...

Page 20: ...bing after sampling by using fingertips to press center ring on adapter while pulling tubing away from the adapter To set up and run a timed sample 1 Set sample duration as follows From Main Menu Touch Sample Touch Flow or Advanced Hour digit 1 will flash Touch up down arrow buttons to increment decrement hour Touch right arrow to advance to hour digit 2 will flash and up down arrow buttons to adj...

Page 21: ...ch Presets Touch desired Preset programmed in DataTrac Pro Software and uploaded to pump Presets are created by the user in DataTrac Pro for Pocket Pump TOUCH Software and uploaded to the pump see DataTrac Pro for Pocket Pump TOUCH User Manual for details on creating presets The start date for a preset with delayed start can be changed on the pump by touching the date on the Preset screen The pump...

Page 22: ...kinked tubing can restrict airflow and cause back pressure to build to a point at which the pump can no longer compensate flow within 5 If this condition is sustained for 3 to 10 seconds the pump will go into flow fault mode as follows 1 Pump stops running and status LEDs on pump flash red Elapsed time stops 2 An exclamation point icon appears on the display next to the flow rate 3 After 20 second...

Page 23: ...o 877 94 and DataTrac Pro for Pocket Pump TOUCH Software which is available as a download requires Internet connection Access DataTrac Pro for Pocket Pump TOUCH Software using the following procedure 1 Check that PC meets DataTrac Pro system requirements Operating system Microsoft Windows 7 8 8 1 or 10 Required software DataTrac Installer included with installation Minimum display resolution 1024 ...

Page 24: ...ttery pack housing with pump case b Use the back of the battery pack housing to push up the belt clip until the battery pack is aligned with the pump case and presses easily into place Ensure the battery pack housing fits flush with the pump case c Use a 1 5 mm hex key to tighten two screws on back of the battery pack housing Tighten screws in an alternating fashion d Charge the battery pack compl...

Page 25: ... 224 26 03 224 26 04 Tube Covers A 6 mm OD x 70 mm L B 8 mm OD x 110 mm L C 10 mm OD x 150 mm L D 10 mm OD x 220 mm L 224 29A 224 29B 224 29C 224 29D Replacement Battery Pack Li Ion P76303 Single Charging Cable USB without wall cube P75739 Low Flow chek mate Calibrator 20 to 500 ml min includes a 9 volt alkaline battery and NIST traceable calibration certificate 375 00205N Replacing baƩeries with ...

Page 26: ... below sea level Flow Fault After 3 to 10 seconds of restricted flow pump stops running elapsed time stops status LEDs flash red and pump displays fault icon After 20 seconds in fault auto restart is attempted up to 5 times unless full airflow is restored prior to that If full airflow is not corrected after 5 auto restarts within 5 minutes the pump ends the run and displays Sample Summary Display ...