53
Table 206 Recommended Separation Distances between portable and mobile RG
Communications equipment and bed.
Equipment and Systems that are NOT Life-Supporting
Recommended Separations Distances for the Bari Rehab Platform 3™: the Bari Rehab Platform
3™ is intended for use in the electromagnetic environment in which radiated disturbances are
controlled. The customer, or user, of the Bari Rehab Platform 3™ can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
Communications Equipment and the bed, as recommended below, according to the maximum
output power of the communications equipment.
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
Compliance Level Cond RF 3 Rad RF-800 MHz 3 Rad RF – 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
Table 208 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Equipment and Systems that are NOT Life-Supporting
The bed is intended for use in the electromagnetic environment specified below. The customer
or user of the bed should ensure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
The bed may be used in shielded or
unshielded locations. Each cable
that enters a shielded location must
meet the minimum FR Filter
requirements.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Field strengths outside the shielded
location from fixed RF transmitters,
as determined by an electromagnetic
site survey, should be less than 3
V/m.