Operating Instruction SIROSONIC/L/TL, PerioSonic
5 Operation
62 78 415 D3275
.
201.04.09.02
27
5
Operation
Use only faultless tips
NOTICE! Tips wear out and can bend resulting in a performance loss
and risk of breakage. Use only ultrasonic tips that are not worn or bent.
Do not reshape bent tips.
Clamping tips
CAUTION! A loose ultrasonic tip can come loose from the thread or
break off. This may cause injury! Use the ultrasonic handpiece only if
the ultrasonic tip is firmly clamped in place.
Overheating at the preparation site _ flow rate
CAUTION! Insufficient cooling leads to overheating of the
preparation site and damage to the tooth substance. Ensure that the
flow rate is > 30 ml/min.
Puncture wound _ Tip
CAUTION! You can puncture your hand with an ultrasonic
handpiece when the tip is inserted. This may cause injury! When the
ultrasonic handpiece is not in use, remove the tip or attach the torque
wrench for protection.
Harmful to eyes
CAUTION! Potentially hazardous optical radiation may cause harm
to the eyes. Do not stare at the LED for longer periods of time while in
operation.