Operating Instruction SIROSONIC/L/TL, PerioSonic
5 Operation
62 78 415 D3275
.
201.04.09.02
31
5.3
Inserting and removing the ultrasonic tip
Inserting the ultrasonic tip
Insert using a torque wrench
✔ The ultrasonic handpiece is not in operation.
1. Insert the ultrasonic tip into the torque wrench (A).
2. Slowly screw the ultrasonic tip into the torque wrench clockwise.
Turn a quarter revolution past the resistance.
3. Detach the torque wrench.
4. Check the ultrasonic tip [ → 32].
Clamping tips
CAUTION! A loose ultrasonic tip can come loose from the thread or
break off. This may cause injury! Use the ultrasonic handpiece only if
the ultrasonic tip is firmly clamped in place.
Removing the ultrasonic tip
Remove the tip with the torque wrench
✔ The ultrasonic handpiece is not in operation.
1. Attach the torque wrench to the ultrasonic tip.
2. Unscrew the ultrasonic tip by turning it counterclockwise.