7 Conditioning
Operating Instruction SIROSONIC/L/TL, PerioSonic
68
62 78 415 D3275
.
201.04.09
.
02
7.2.2
... with cleaning and disinfection equipment
No strong acidic, alkaline or chlorine solutions
NOTICE! Do not use any strong acidic or alkaline solutions (5 < pH < 9)
or solutions containing chlorine.
Scaler
The SIROSONIC/L/TL / PerioSonic, torque wrench and ultrasonic tips
can also be cleaned and disinfected in suitable cleaning and
disinfection equipment.
NOTICE! The ultrasonic tips can be cleaned and disinfected on the
outside only after removing them from the torque wrench.
Cleaning and disinfection equipment
The cleaning and disinfection equipment used must be approved by its
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
Checking without maintenance _ Instruments (plural)
✔ The instruments are conditioned with a cleaning and disinfection
device.
6278415.book Seite 68 Montag, 17. Oktober 2016 11:28 11