Maintenance
Never oil the scaler
NOTICE! Never oil the ultrasonic handpiece!
8.1
Maintaining the water channel
Scaler without SiroAir L
CAUTION! Insufficient cooling leads to overheating of the
preparation site, the tip, and the handpiece. Ensure that the water
content is > 30ml/min.
Exception: The water flow is blocked in tip no. 6L (CEM). When using
tip no. 6L (CEM), the ultrasonic handpiece can be operated only
intermittently.
1. Use the cleaning wire to carefully clear the water channel of the tip
regularly.
2. Blow the water channel out with max. 3 bar.