Operating Instruction SIROSONIC/L/TL, PerioSonic
7 Conditioning
62 78 415 D3275
.
201.04.09.02
69
1. Check whether the instruments are clean after conditioning under
good lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If they are still dirty, repeat the process.
ª
For further processing, the instruments should be dry and free of
residue.
3. Blow the instruments out with max. 3 bar.
4. Pack the instruments in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
5. Conduct sterilization [ → 72].