6 Follow-up
Operating Instruction SIROSONIC/L/TL, PerioSonic
64
62 78 415 D3275
.
201.04.09
.
02
6
Follow-up
6.1
After each treatment session
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
Scaler
Puncture wound _ Tip
CAUTION! You can puncture your hand with an ultrasonic
handpiece when the tip is inserted. This may cause injury! When the
ultrasonic handpiece is not in use, remove the tip or attach the torque
wrench for protection.
✔ The ultrasonic handpiece is not in operation.
✔ Wear appropriate protective clothing.
1. Purge the water lines directly at the treatment unit for 30 seconds.
2. Predisinfect directly at the treatment center [ → 66].
3. Unscrew the ultrasonic tip using the torque wrench.
4. Detach the ultrasonic handpiece.
6278415.book Seite 64 Montag, 17. Oktober 2016 11:28 11