3 Safety
Sirona Dental Systems GmbH
3.1 Basic safety information
Operating Instructions Sirona Dental CAD/CAM System
3
Safety
Safety
3.1
Basic safety information
Basic safety information
3.1.1
Prerequisites
Prerequisites
3.1.2
Connecting the unit
Connecting the unit
Description
Perform connection by following the directions given in the present
operating instructions.
3.1.2.1
Connection to the CEREC 3 acquisition unit
Connection to the CEREC 3 acquisition unit
Warning: inEos connected to CEREC 3 acquisition unit
NOTICE
Important information on the building installation
The building installation must be performed by a qualified expert in
compliance with the national regulations. DIN VDE 0100-710 applies in
Germany.
NOTICE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion
hazards.
NOTICE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
WARNING
No acquisitions in the patient's mouth
As long as an inEos Blue is connected to a CEREC 3 acquisition unit,
no acquisitions may be made in the patient's mouth using the connected
CEREC 3D camera.