10 Calibration
Sirona Dental Systems GmbH
10.3 Calibrating the XYZ axis alignment
Operating Instructions Sirona Dental CAD/CAM System
ª
The window for calibrating the 3D camera system opens.
ª
You will be prompted to attach the calibration part.
2. Fasten the calibration part to the left side of the cone with the clip.
3. Tilt the calibration part via the cone.
4. Slide the calibration part upward up to the stop.
5. Confirm the message box with the
"OK" button.
ª
The calibration is performed.
ª
A message appears stating that the unit has been successfully
calibrated.
6. Confirm the message box with the
"OK" button.
7. Remove the calibration part.
Note: After each additional calibration of the camera system, you also
must calibrate the X-Y-Z axis alignment.
10.3
Calibrating the XYZ axis alignment
Calibrating the XYZ axis alignment
1. Click
"Settings"/"Calibration"/"inEos"/"XYZ alignment".
Tip: The following dialog boxes will guide you through the calibration
procedure and provide all relevant information.
ª
The calibration window and a message box open.
2. Check to make sure that the calibration parts are not damaged and
click the
"OK" button .
Step 1
Step 1
1. Place the calibration plate in position 1.
Note the position of the beveled corner. The grid balls must be seated
as shown in the illustration.
The right half of the cylinder is located in the image field.