Calibration
Operating Instructions Sirona Dental CAD/CAM System
10
Calibration
Calibration
10.1
General information on calibration
General information on calibration
The inEos Blue has two components that must be calibrated:
● 3D camera system
● XYZ axis alignment
Various calibration standards and parts are required for calibration (see
Chapter on "Calibration parts" [ → 16]). They are included in the scope of
supply of the inEos Blue.
Recalibrate your inEos Blue
● following transport (shaking stress),
● after storage in unheated or un-air-conditioned rooms (temperature
differences exceeding 30°C),
● with temperature differences of over 15°C between the last
calibration and operation.
10.2
Calibrating the 3D camera system
Calibrating the 3D camera system
Preparing the inEos Blue
✔
The PC and the inEos Blue are switched on.
✔
The inLab software is started
1. Click
"Settings"/"Calibration"/"inEos"/"Camera system".
NOTICE
Calibrate prior to initial startup
Prior to initial startup of the inEos Blue, you must perform a full
calibration. No further calibrations are usually required afterwards.
If you remove the shifting plate, the X-Y-Z alignment must be
recalibrated after reassembling the unit.
NOTICE
Do not damage the calibration parts
Proceed carefully when attaching or removing the calibration part in
order to ensure that it is not damaged or contaminated.