13 List of error codes
Sirona Dental Systems GmbH
11.7 Exposure times using tube with 12" SSD - General X-ray sensors (Round or square cone)
Operating Instructions HELIODENT DS
60 81 744 D 3302
26
D 3302.201.01.25.02
13
List of error codes
Errors occurring during the self-test are indicated by a numeric error code of 1 - 4.
CAUTION
If an error is displayed again after switching the unit off
and then on again, please contact your service techni-
cian.
Error number
Description of cause
Corrective action
No. 1
The cable connecting the X-ray tube unit
to the generator may be interrupted (kV
line).
Switch unit OFF.
Contact your service technician.
No. 2
X-ray was not exposed
(mA/kV
max
error).
Switch unit OFF.
Contact your service technician.
No. 3
D
button was pressed during switch-on
phase.
Switch unit OFF and ON again.
Be sure not to press the D button while doing so.
No. 4
Release button was pressed during
switch-on phase.
Switch unit OFF and ON again.
Be sure not to press the release button while doing so.