5 Preparing an exposure
Operating Instructions HELIODENT DS
5.2 Selecting the radiographic technique
60 81 744 D 3302
D 3302.201.01.25.02
13
båÖäáëÜ
5.2
Selecting the radiographic technique
Digital
If you are working with a digital radiographic system (e.g.
SIDEXIS
), the
Digital
mode indicator lamp (6) on the
unit
must
be lit up.
You can switch back and forth between the conventional
and the digital radiographic technique by briefly holding
down the
D
button. The resulting reduction in the expo-
sure time will then be displayed.
Use an X-ray field limitation for the digital radiographic
technique.
NOTE
i
To set the recommended exposure times for adults, al-
ways set the patient symbol to the highest exposure level
within the relevant tooth symbol.
If you are working with children, always set to the lowest
exposure level within the relevant tooth symbol!
Conventional
If you want to take conventional (film) exposures, the
Digital
mode indicator (6) must
not
be lit up.
You can switch back and forth between the conventional
and digital radiographic technique by briefly pressing the
D button. The
Digital
mode indicator (6) then switches
off.
The exposure time for conventional exposures is then
displayed.
Insert the X-ray field limitation for the conventional radi-
ographic technique.
0.06
D
S
X-ray
Digital
7 mA 60 kV DC
6
0.25
D
S
X-ray
Digital
7 mA 60 kV DC
Summary of Contents for HELIODENT DS
Page 2: ......
Page 4: ......
Page 5: ...aÉìíëÅÜ 0 06 D S X ray Digital 7 mA 60 kV DC HELIODENT DS ebiflabkq ap dÉÄê ìÅÜë åïÉáëìåÖ ...
Page 33: ...båÖäáëÜ ebiflabkq ap léÉê íáåÖ fåëíêìÅíáçåë 0 06 D S X ray Digital 7 mA 60 kV DC HELIODENT DS ...
Page 61: ...cê å áë ebiflabkq ap kçíáÅÉ ÇÛìíáäáë íáçå 0 06 D S X ray Digital 7 mA 60 kV DC HELIODENT DS ...
Page 89: ...bëé çä ebiflabkq ap fåëíêìÅÅáçåÉë ÇÉ ìëç 0 06 D S X ray Digital 7 mA 60 kV DC HELIODENT DS ...
Page 117: ......