Sirona Dental Systems GmbH
6 Positioning the patient/x-ray tube unit
Operating Instructions HELIODENT DS
5.5 Setting the exposure data
60 81 744 D 3302
D 3302.201.01.25.02
17
båÖäáëÜ
6
Positioning the patient/x-ray tube unit
Adjust the X-ray tube unit
Have the patient sit down on the chair.
Right-angle technique
(with X-ray field limitation)
The film or X-ray sensor is positioned with a holding sys-
tem.
Sirona X-ray sensors may be used only with the holding
systems recommended by us.
Always follow the Operating Instructions for intraoral
radiography supplied with SIDEXIS.
45° angle technique (without X-ray field
limitation)
The film or X-ray sensor must be held by the patient.
Angle of inclination
Exposure tube for the occlusal plane
Supramaxillary teeth
Molars
35
×
Bicuspids and canine teeth
45
×
Front teeth
55
×
Occlusal exposure
80
×
Occlusal exposure
-90
×
Front teeth
-20
×
Bicuspids and canine teeth
-10
×
Molars
-5
×
Submaxillary teeth