SensorIQEasy_UM_9752023110_NL_D_GB_V2_0
57
GB
6.5 Setting the alarm set point (expert mode)
The Sensor IQ Easy bar will generate a warning if the average measured value reaches the
Alarm set point (peak)
(during
warning/error delay
see 6.6)
.
This value is set to 15kV by default, but it can be easily adjusted:
- [
(n)x
“
Alarm set point (peak)
”
slide slider or use
to go to the desired value and confirm with
].
6.6 Setting warning/alarm error delay (expert mode)
The Sensor IQ Easy bar will generate a warning if the average measured value continually
exceeds the
set point
during
warning/error delay.
This value is set to 10 seconds by default, but it can be easily adjusted (minimum value 2 sec.):
- [
(n)x
“
Alarm set point (peak)
”
enter the desired value and confirm with
].
6.7 Sensor Standby & Active
By default, the bar will become active after all information has been gathered by the manager.
You can set the device in standby
or active
mode on each tab.
The status is indicated in the left upper corner by means of a colour behind the
icon:
Green = standby, blue = running, orange = warning, red = alarm.
6.8 Setting bar information parameters (expert mode)
If desired, various information parameters can be entered as required to make the different
devices more recognisable.
If desired, adjust the parameters: Device name, Machine position,
These are information parameters neither the bar nor the manager will use for calculations.
- Select the information page with the parameters to be changed by:
- [
,
,
(n)x
“parameter
“, enter name or value”, confirm with
]