5.3 Placing Sensor IQ Easy
Note:
- Conductive machine parts close to an IQ Easy Sensor may have a negative influence
on its effectiveness. For the best results the Sensor should be mounted according to
figure 7.
- Mount the Sensor IQ Easy:
•
With the sensor(s) in the direction of the material to be measured.
•
Behind the anti-static bar. At least 2x the operating distance of the anti-static bar.
•
at 10 mm (0-1000 V), to 50 mm (0-20 kV), to 100 mm (0-30kV), to 300 mm (0-80 kV)
distance from the material to be measured.
(mounting distance should be entered into the parameters via the Manager IQ Easy see
H6.3)
* For Segment units with external sensors, observe the same distances for the sensors
with respect to the material to be measured. The position of the rod relative to the
measuring surface is not important.
Figure 6: Position Sensor IQ Easy towards the anti-static bar