Instruction Manual OilGuard 2 W
Electrical installation
32
14809E/2
5.5.2
Connecting the Modbus RTU or Profibus DP
The terminals on the Profibus DP or Modbus RTU module are assigned as follows:
TERMINALS
MODBUS / PROFIBUS
FUNCTIONAL DESCRIPTION
11
Ground IN
Connection for cable shielding
12 A
RS 485-A IN
Data connection
13 B
RS 485-B IN
Data connection
14
Ground OUT
Connection for cable shielding
15 A
RS 485-A OUT
Data connection
16 B
RS 485-B OUT
Data connection
5.5.3
Overview of Profinet IO
To connect to the Profinet IO, the Profinet IO module must be integrated in the Oil-
Guard 2 W.
The module has an internal switch and provides two Ethernet ports.
The cable is connected directly to the RJ45 plug of the Profinet IO module inside the in-
strument or via external M12 connectors.
When connecting directly to the RJ45 plug, please note that only plugs with a short
and flat design can be used.
In the
Digi.interf.\General
menu, the
Modul type
must be set to
Profinet IO
.
In the
Digi.interf.\Profinet
menu, the station name, MAC address and connection sta-
tus are shown. Moreover, it can be defined here whether the data should only be read
or be read and written.
Figure 7: Overview of the Profinet IO module
Field bus interface for Profinet IO
(connection print)
Connector to the AQ2Basi print
Ethernet port 1
Ethernet port 2