Servicing
Instruction Manual OilGuard 2 W
14809E/2
65
9.1.5
Cleaning parts which come into contact with water
The following describes cleaning the parts of the OilGuard 2 W that come into contact with
water:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the sample supply to the photome-
ter.
Section 4.4
2.
Remove the inlet and outlet from the pho-
tometer.
3.
Remove the optics unit on the photometer
according to Section 9.1.1 and fasten on the
docking station.
Danger UV radiation.
Use UV goggles and gloves.
4.
Loosen the conduit gland (A) and remove the
inlet pipe (B).
5.
Remove the outlet cone (C) by pulling it up-
wards.