Instruction Manual OilGuard 2 W
Servicing
64
14809E/2
9.1.4
Automatic adjustment with OilGuard 2 W A
Automatic adjustment is only possible with the instrument type OilGuard 2 W A. The au-
tomatic adjustment can also be set for time actuation in the
Adjust interval
menu. This is
described in the Reference Manual.
The following procedure describes how manual triggering of an adjustment is made with an
OilGuard 2 W A:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Switch the photometer to service operation.
Section 7.10
2.
Press the
Recalibration
button and then
C1
EPA-PAH
.
3.
Carry out the adjustment as follows:
Press the
initiate...
button. This starts the ad-
justment.
If the adjustment was successful, this is con-
firmed with
Adjustment OK
. This completes
the adjustment.
If the calibration check was
not
successful, it is
indicated with
Adjust. fault
.
A possible cause of this is soiled optics in the
instrument.
If the check could not be suc-
cessfully completed, contact your
country representative (Section 11).
A new recalibration factor is determined during the adjustment. The deviation from the
original state is displayed under
Curr. corr.
.