Instruction Manual DualScat Ex
10118E/5
21
Meas. range No. Meas. span
(standard)
Meas. span
(user-specific)
1
0 .. 2000 NTU
2
0 .. 500 NTU
3
0 .. 100 NTU
4
0 .. 50 NTU
5
0 .. 20 NTU
6
0 .. 10 NTU
7
0 .. 5 NTU
8
0 .. 2 NTU
MR inputs
Remote range switching
Reference Handbook
Automatic
0 .. 2000 NTU
Table 3: Measuring ranges
If you need different measuring ranges, have a SIGRIST service technician
adapt these and enter new ranges in "user-specific" column.
4.7
Setting the relay functions
The control unit has two relay outputs (
Section 3.2), the functions of which
are freely configurable. Several functions can be assigned to a given relay sim-
ultaneously. The relay in question is activated whenever
one
of the configured
functions becomes active (OR link).
Action
Display (Example)
Remarks
1.
Access code
> ▓00000 <
If no individual access
code was set, proceed to
step 3.
2.
Enter code:
/
change number
/
change place
Access code
> ...... <
Enter your own access
code here.
3.
* SENSOR *
* CHECK *
Instrument in service
mode.
4.
4 x
* CONFIGURATION*
5.
8 x
> Relay 1 <
li al se ch in
Configure relay 1.
6.
Relay 1
>li al se ch in<
Activate editing mode.
Relay functions