Instruction Manual DualScat Ex
20
10118E/5
4.6
Setting the measuring range
On the dual-angle instrument, the measuring ranges for the 90° and 25°
measurements can be set separately. If "Automatic" is set, the photometer
switches automatically to the optimum measuring range for the momentary
measurement.
Action
Display (Example)
Remarks
1.
Access code
> ▓00000 <
If no individual access
code was set, proceed to
step 3.
2.
Enter code:
/
change number
/
change place
Access code
> ...... <
Enter your own access
code here.
3.
* SENSOR *
* CHECK *
Instrument in service
mode.
4.
2 x
* MEAS.RANGE *
* LIMITS *
5.
Select meas. angle:
(1
st
meas. angle)
(2
nd
meas. angle)
> Meas.range xx<
Automatic
6.
Meas.range xx
> Automatic <
Activate editing mode.
7.
Select meas. range:
/
Meas.range xx
> ... <
Table 3 below.
8.
> Meas.range xx<
...
Confirm selection.
9.
+
(simultaneous)
90° 2.23 NTU 5
Instrument in normal op-
eration.
Measuring range