Instruction Manual DualScat Ex
14
10118E/5
Terminals
Used for
Remarks
1.
20 .. 23
Connection to photometer
SIREL SMD
: Connect
wires to the terminal
block according to color
coding.
SIREL Ex
: Connect wires
according to numbering.
2.
4 - 5 - 6
Relay output 1
The relay outputs are
freely configurable.
Reference Handbook
7 - 8 - 9
Relay output 2
18 - 19
Reading output for
1
st
meas. angle
(90° reading in the case of
dual-angle instrument)
0/4 .. 20 mA, max. bur-
den 600
If unused, these terminals
must be shorted with a
jumper!
32 - 33
Reading output for
2
nd
meas. angle (25° read-
ing in the case of dual-
angle instrument)
Provided on dual-angle
instrument only.
If unused, these terminals
must be shorted with a
jumper!
3.
10 .. 17
Control signals for
1
st
meas. angle
Reference Handbook
26 .. 31
Control signals for
2
nd
meas. angle
Provided on dual-angle
instrument only.
16a
Control signal for sensor
check
Reference Handbook
24 - 25
Connection to bus coupler
Reference Handbook
4.
1 - 2 - 3
Mains voltage
100 .. 240 V,
47 .. 63 Hz or 18 ..
30 VDC, 25 W
Connection